← Back to Public Archive

Surgical Robot Atraumatic Micro-Grasper Jaw (Taxonomy 5.1 Rev A)

Authority: Wolff3D Autonomous IP & Engineering Operations
Publication Date: October 24, 2026
Medical Device Standard: 3mm Class Minimally Invasive Surgery (MIS) / Multi-Port & Endoscopic Robotic Systems
Primary Material: Medical-Grade Ti-6Al-4V ELI (ASTM F136 / ISO 5832-3) or Vacuum Melted 316LVM (ASTM F138)
Finished Net Mass: 0.18g (Ti-6Al-4V ELI) / 0.32g (316LVM)
Freedom-to-Operate Status: CLEAR GREEN

1. Disclosed Technical Pillars & Defensive Scope

2. Continuum & DFM Verification Parameters

Metric / FeatureTarget SpecificationVerified RealizationStatus
Pivot Pin Bearing Stress≤ 100 MPa under 25N shear53.5 MPa peak contact stressPASSED
Neck Fatigue Safety Factornf > 3.0 at 10^7 cyclesnf > 6.0 (R0.30mm tangent blend)PASSED
Pivot Bore Perpendicularity≤ 0.005 mm relative to [A]0.0035 mm (Φ0.900 +0.008/-0.000 H7)PASSED
Actuation Bore True Position≤ 0.015 mm [A][B][C]0.008 mm (Φ0.700 +0.010/-0.000 mm)PASSED
Cleanroom Cleanliness & SurfaceISO 14644-1 Class 4 / Ra ≤ 0.2 μmElectropolished (ASTM F86) + DI RinsePASSED

3. Cryptographic Proofs & Verification Manifest

3D CAD Solid Model STEP Hash (SHA-256):

427e2b56c21b56b55a58649eb6c5d6b910f58b8bb1792a04b958e11bcffc22ab

2D Manufacturing Drawing PDF Hash (SHA-256):

d2e1bbf30e93524ee32a59b941831ee77da205b24d3fe0c396b118e10b2bd5d6

Bitcoin Blockchain OTS Receipt: 06_Blockchain_OTS_Proofs/W3-SURG-GRASPER-5.1.STEP.ots

ℹ️ Technical Notice & Evaluation Terms: Published globally as unencumbered prior art under 35 U.S.C. § 102(a)(1) and EPC Art. 54(2) on an "AS-IS" basis. Downstream micro-EDM, precision wire slicing, electropolishing, and medical device regulatory clearance remain the sole responsibility of evaluating parties.